Model No. |
PGC- 1 |
PGC- 2 |
PGC- 3 |
PGC- 4 |
PGC- 5 |
Usable Space in MM |
610 x 450 x 410 |
845 x 500 x 400 |
870 x 570 x 560 |
900 x 650 x 570 |
700 x 900 x 650 |
Capacity |
4 Cu.ft. |
6 cu. ft |
10 cu. ft |
12 cu. ft |
15 cu. ft |
Approx Volume (Ltr) |
113 ltrs |
170 ltrs |
284 ltrs |
340 ltrs |
425 ltrs. |
No. of Shelves |
2 |
2 |
2 |
3 |
3 |
Temperature Range |
5°C above ambient to 60°C
(Resolution 0.1°C) |
Temp. Controller |
By Microprocessor Based PID Digital
Temperature/Humidity Indicator cum Controller. |
Temp. Accuracy |
+1°C |
Humidification |
By steam injection process. |
Humidity Range |
5% above ambient from 40% to 95% RH at
cool temperatures. |
Humidity Accuracy |
+ 5% RH |
Display: |
LED Display for Set Value(SV) and
Process Value (PV) |
Relay |
Solid state electronic relay with
protective heat sink. |
Air Circulation |
By forced convection system |
Safety |
By safety thermostat to prevent
overheating |
Illumination |
Exterior illumination with fluorescent
tubes/lamps. |
Insulation |
By High density PUF insulation |
Operations |
Nearly silent operation with ultra-low
vibration |
Voltage Indicator |
By Digital Volt Meter fitted on Panel |
Electric Supply |
220/230V AC, 50/60Hz |
Controlling of Temperature |
By Profile Digital Microcontroller
having 4 Programmers each of16 steps (total 64 steps of ramp / soak
profile) |
PLC Based Control Systems |
With Touch Screen Display with Direct Thermal / Dot
Matrix Printer. |
LCD Controller with Data Logger |
(16 x 2) with optional RS-485 communication ports,
cables, window based software with inbuilt data recording. For
(Temperature, humidity and light levels) storage up to 2500 events. |
Communication Port |
Communication Port with interface and data cable to
download data to your PC. |
Lighting Control |
Five Levels light control (0%, 25%, 50%, 75% and
100%) |
Timer |
With reverse 0 - 24 hours for regulating hours of
light and darkness |
Real Time Clock |
Facility can be incorporated |
Alarm Facility |
Audio/Visual alarm for temperature deviation from
set values. |
High Voltage Protection |