Home »
Our Products »
Clean Air » Laminar Air Flow
(Horizontal)
Laminar Air Flow (Horizontal)
APPLICATIONS
- "Square" Laminar Air Flow Benches Horizontal are designed
so as to meet the requirement of US Federal Standard 209 B (BS 5295)
providing particle free air to meet class 100 conditions mainly used for
various works that require maintenance of the cleanness and purity.
- It is Ideal for Culture analysis, Bacteriological & Biological
applications in Hospitals, Plant Tissue Culture, Labs, Food, Beverage
and Electronic industries.
- It provides an economical work environment- suitable for hospitals,
drugstore, pharmaceuticals factories, sanitation and anti-epidemic
stations.
SALIENT FEATURES &
CONSTRUCTION
- Low Cost.
- Low Noise.
- User Friendly Ergonomic Design.
- Controlled Environment Particulate Free Product Protection.
- Basic material is thick board wood front back top and exterior
surfaces are covered with Formica. OR Cold rolled steel duly powder
coated OR Stainless Steel (SS-304) (as ordered).
- Working table is made of Stainless Steel Sheet (SS-304 grade).
- All interior surfaces are epoxy painted filter seat is carefully
formed, installed and checked for leakage.
- Side panels & front doors are made of 6 mm thick acrylic
transparent plexi glass.
- The highly efficient HEPA filter maintains the optimum cleanliness
and purity.
- The pre-filter fitted extends the life of HEPA filter drastically.
Air drawn through pre-filter is made to pass through highly effective
HEPA (High Efficiency Particular Air) filters having efficiency rating
as high as 99.99% with cold DOP and 99.97% with hot DOP, thus retaining
all air-borne particles of size 0.3 micron and larger.
- Working area is illuminated by fluroescent lighting fitted to the
unit.
- The blower and motor assembly is statically and dynamically balanced
of 1/5 H.P. capacity operates with minimum noise level.
- Height of the working table provides comfortable "SIT DOWN"
working position for the operator.
- Supplied complete with S.S. Table Top, Transparent front door, Static
Pressure Manometer. Built in U.V. Germicidal Light and Cock for gas, air
or vacuum line, HEPA filters and pre-filter.
- Supplied complete with cord & plug to work on Single Phase 220 /
230 volts A.C. supply.
TECHNICAL SPECIFICATIONS & ORDERING INFORMATIONS
Model No |
HLF-1 |
HLF-2 |
HLF-3 |
HLF-4 |
Working Table Size |
2' x 2' |
3' x 2' |
4' x 2' |
6' x 2' |
Working Size |
2' x 2' x 2' |
3' x 2' x 2' |
4' x 2' x 2' |
6' x 2' x 2' |
HEPA Filter Size |
2' x 2' x 6” |
3' x 2' x 6 |
4' x 2' x 6” |
3' x 2' x 6” |
No. of Motor/Blower |
1 |
1 |
1 |
2 |
Illumination |
1 x 20 w |
1 x 20 w |
2 x 40 w |
2 x 40 w |
Main Filter |
HEPA Filter (99.97% efficient at 0.3
micron particles) |
Pre-Filter |
High efficiency pre-filter (Washable) |
Sterilization |
By U.V. Germicidal Tube in work area |
Front Door |
Made of Plexi Glass 6 mm |
Noise Level |
Less than 65 db |
Standard Accessory |
Gas/Air cock |
Electric Supply |
AC 230V , 50/60Hz |
OPTIONAL FEATURES
Spare HEPA Filters |
Can be provided at an extra cost |
Magnetic Gauge |
(To track filter pressure) in lieu of static
pressure manometer. |
Interlocking |
U.V . Germicidal tube & fluorescent light with
auto switch off/on mode for motor blower when the door is closed/open. |
Hour Meter totalizer |
For U.V . light |
Microprocessor LCD Controller with Auto Switch off
mode |
For motor blower when door is closed. Digital buzzer
timer sounds an audio alarm on completion of sterilizing time of the
chamber and the switches off U.V . germicidal tube and automatically
switches ON the fluorescent light, alerting the user that the bench is
ready for use |
Can be fitted with micro chip controlled electronic
digital indicator |
T o display air flow rate fitted with sensor |